Anesthesia Workstation for newborns, children and adults with integrated monitor to display the respiratory parameters of the patient. It is a continuous pneumatic, electronically controlled anesthesia ventilator with an integrated tube for inhalation anesthesia with oxygen, nitrous oxide and two anesthetic gases.
| Parameter | Practice 3700 | Practice 3700 (electronic flowmeters |
|---|---|---|
| Integrated fan | Power: from the gas supplyVentilation modes: IPPV, PCV, SIMV, PS, ManualTidal Volume: 20 – 1500 mlVentilation frequency: 2 – 100 bpmRatio I / E: 4:1 – 1:8Pressure range: 5 – 70 cm H2OPEEP level: 3 – 30 cm H2Oinspiratory plateau: off, 5% – 60% TiTrigger sensitivity (flow): 1 – 15 l / min | Power: from the gas supplyVentilation modes: IPPV, PCV, SIMV, PS, ManualTidal Volume: 20 – 1500 mlVentilation frequency: 2 – 100 bpmRatio I / E: 4:1 – 1:8Pressure range: 5 – 70 cm H2OPEEP level: 3 – 30 cm H2Oinspiratory plateau: off, 5% – 60% TiTrigger sensitivity (flow): 1 – 15 l / min |
| Monitoring | 10.4 inch TFT color screenGraph: P-T, F-TCurves: P-V, F-V O2 concentrationPressure (Ppeak, Pplat, Pmean, PEEP)respiratory rateVolume: MV, VtBattery Charge Indicator | 12.1 inch TFT color touch screenGraph: P-T, F-TCurves: P-V, F-V O2 concentrationPressure (Ppeak, Pplat, Pmean, PEEP)respiratory rateVolume: MV, VtBattery Charge Indicator |
| Respiratory system | closed, semi-closed, semi-open | closed, semi-closed, semi-open |
| Range of fresh gas flow indicators | Oxygen: 0-1, 1-10 l / min N2O: 0-1, 1-10 l / min Air: 0-1, 1-12 l / min |
O2, N2O : 0.05 – 10.0 l/min Air: 0.05 – 12.0 l/min O2 flush: 25 – 75l/min |
| Anesthetic gases | Halothane, Enflurane, Isoflurane, Sevoflurane available | Halothane, Enflurane, Isoflurane, Sevoflurane available |
| Alarm | Paw high/low limitMV high/low limitRespiratory rate is too highFiO2 high/low limitApneahigh PEEP levelPower failure | Paw high/low limitMV high/low limitRespiratory rate is too highFiO2 high/low limitApneahigh PEEP levelPower failure |
| Gas (O2, N2O, air) | 0.28-0.6 МPa | 0.28-0.6 МPa |
| Electric supply | 100 – 240V AC, 50/60 Hz, built-in rechargeable battery 24V DC (min. 120 min) | 100 – 240V AC, 50/60 Hz, built-in rechargeable battery 24V DC (min. 120 min) |